Knowledge Battery Testing Which specific parameters must laboratory battery testing instruments measure to accurately evaluate the State of Health (SoH) of single cells and cell strings?
Author avatar

Tech Team · Kintek Solution

Updated 3 months ago

Which specific parameters must laboratory battery testing instruments measure to accurately evaluate the State of Health (SoH) of single cells and cell strings?


The essential SoH measurements are deliverable capacity, internal resistance, and self-discharge current. For a single cell, laboratory equipment should measure these directly and record cell voltage, charge/discharge current, and temperature throughout the test. For a cell string, it must additionally resolve capacity imbalance, cumulative string resistance, and the self-discharge behavior of the weakest cell.

SoH is determined primarily by how much capacity a cell can still deliver, how much its internal resistance has increased, and how quickly it loses charge while at rest. In a string, the weakest cell and the degree of imbalance can determine practical health even when the string’s average values appear acceptable.

What a Laboratory Instrument Must Measure

Actual deliverable capacity

The instrument must measure the usable ampere-hour capacity obtained during a controlled discharge, not merely the manufacturer’s nominal rating.

Capacity is determined by integrating discharge current over time under defined conditions. The test should specify the charge method, discharge current or C-rate, voltage limits, temperature, rest periods, and cutoff criteria so results can be compared with the cell’s fresh baseline.

Capacity retention is commonly expressed as:

[ \text{Capacity-based SoH} = \frac{C_{\text{current}}}{C_{\text{initial}}} \times 100% ]

A declining capacity indicates loss of active material, increased polarization, reduced lithium inventory, or other degradation mechanisms.

Internal resistance

The instrument must measure internal resistance or dynamic impedance, preferably using a defined test method and operating condition.

Resistance can be evaluated through a current step, pulse test, AC impedance measurement, or electrochemical impedance spectroscopy. The selected method matters because measured resistance changes with state of charge, temperature, frequency, current amplitude, and measurement duration.

A useful resistance-based indicator is:

[ \text{Resistance-based SoH} = \frac{R_{\text{initial}}}{R_{\text{current}}} \times 100% ]

As resistance rises, voltage sag and heat generation increase, reducing the cell’s available power even if its measured capacity remains relatively high.

Self-discharge current or rate

The instrument must measure charge loss during a controlled rest period and convert it into a self-discharge current or rate.

This requires accurate voltage and current measurements, stable temperature control, and sufficient observation time. Self-discharge should be distinguished from measurement leakage, balancing currents, connected electronics, and temperature-driven voltage changes.

An abnormal increase in self-discharge is an important warning sign because it can indicate internal leakage, contamination, separator damage, internal micro-shorts, or other latent faults.

Supporting Measurements Required for Reliable SoH

The three primary indicators cannot be interpreted correctly without recording the operating conditions under which they were obtained.

Cell voltage

The system should continuously record cell voltage, including charge, discharge, rest, and transient behavior.

Voltage data establishes the test limits, reveals voltage sag under load, identifies abnormal recovery after current removal, and allows researchers to compare cells within a string.

For strings, individual cell voltage measurement is essential. Measuring only total string voltage can conceal one weak or overcharged cell.

Charge and discharge current

The instrument must measure current accurately in both directions and integrate it for ampere-hour counting.

Current measurement supports capacity calculation, detects charge or discharge irregularities, and provides the input required for resistance calculations during current pulses. It also allows test results to be normalized across different C-rates.

Temperature

The system should record cell or surface temperature continuously, together with ambient or chamber temperature where relevant.

Temperature strongly affects capacity, resistance, self-discharge, and voltage. Without temperature data, an apparent SoH change may simply reflect a different test temperature rather than permanent degradation.

Cycle count and operating history

A meaningful SoH assessment requires logging cycle count, depth of discharge, C-rate, cutoff-voltage exposure, rest duration, and thermal history.

The same cell can age differently depending on how long it remains at high voltage, how often it operates at high current, and what temperatures it experiences. These stress variables help link measured degradation to a specific failure mode.

Electrochemical impedance

Impedance spectroscopy can supplement simpler resistance measurements by showing how the cell’s response varies with frequency.

It is particularly useful for research because changes in impedance components can help distinguish charge-transfer, diffusion, and ohmic contributions. However, impedance spectroscopy does not replace capacity and self-discharge testing when the objective is practical SoH determination.

What Must Be Measured in Cell Strings

Capacity imbalance between cells

A string test must determine the capacity of individual cells or the effective capacity of each series position, rather than relying only on total string capacity.

The usable capacity of a series string is constrained by the cell that reaches its voltage limit first. A string may therefore have acceptable average capacity while one weak cell limits the usable energy.

Cumulative string resistance

The instrument should measure total string resistance and, where possible, identify the resistance contribution of individual cells and interconnects.

String resistance includes cell resistance, busbars, connectors, fuses, welds, and other conductive paths. A high total value may reflect either cell aging or an external connection problem, so localization is important.

Weakest-cell self-discharge

For a string, the relevant self-discharge assessment must identify the cell with the highest self-discharge rate, not merely calculate an average.

One leaky cell can repeatedly drift out of balance and force the battery management system to reduce usable capacity or initiate balancing. Its behavior may dominate the string’s practical reliability.

Individual cell voltage spread

The system should record minimum, maximum, and individual cell voltages during operation and rest.

Voltage spread under load can reveal resistance differences, while voltage divergence during rest can indicate capacity imbalance or self-discharge. These measurements also verify whether balancing controls are working as intended.

How Instruments Should Capture the Data

Use synchronized, high-resolution channels

Voltage, current, temperature, and timing data should be captured synchronously across all relevant channels.

This is necessary for calculating instantaneous resistance, integrating capacity, comparing cells under identical conditions, and identifying transient events that may be missed by slow or unsynchronized logging.

Control the test conditions

The equipment should provide controlled constant-current and constant-voltage charging, programmable discharge profiles, defined rest periods, and thermal control or compensation.

SoH values are only comparable when the test protocol is repeatable. A capacity result taken at a different temperature or discharge rate is not necessarily evidence of actual degradation.

Validate measurement quality

Before relying on the results, evaluate measurement uncertainty, repeatability, reproducibility, and linearity over the intended operating range.

Illustrative system-level tolerances sometimes used in BMS-related testing include total-voltage error within ±1% of full scale, current error of ±0.3 A up to 30 A or ±1% above 30 A, temperature error within ±2 °C, and module-voltage error within ±0.5% of full scale. These are application-specific requirements, not universal SoH standards.

Record data for degradation analysis

A test system should retain both raw measurements and derived metrics, including capacity retention, resistance growth, self-discharge rate, temperature exposure, cycle count, and cell-to-cell variation.

Historical data allows researchers to identify whether degradation is gradual, accelerated, localized, or associated with particular operating stresses.

Understanding the Trade-offs

Capacity alone is insufficient

Capacity is the most direct measure of remaining energy, but it may not reveal a cell that has developed excessive resistance or abnormal self-discharge.

A cell can still deliver near its rated capacity while exhibiting severe voltage sag, heating, or charge loss during rest.

Resistance alone is insufficient

Resistance is sensitive to temperature, state of charge, test frequency, and pulse duration.

A resistance increase is valuable evidence of degradation, but it must be interpreted using a consistent method and supported by capacity and thermal data.

Average string values can hide weak cells

Total string capacity, average voltage, and average resistance can obscure cell-level failures.

Series strings are limited by their weakest position, so individual-cell measurements are more informative than aggregate values whenever the instrumentation permits them.

Fixed end-of-life thresholds are not universal

An 80% capacity-retention threshold or a doubling of internal resistance is often used as a practical reference, but the correct end-of-life criterion depends on the application, chemistry, safety requirements, and power or energy demands.

For that reason, laboratory testing should establish application-specific thresholds rather than treating one number as universally definitive.

Making the Right Choice for Your Goal

Use the following measurement priorities when specifying a laboratory battery test system:

  • If your primary focus is single-cell energy retention: Require controlled charge/discharge testing with accurate current integration, cell-voltage measurement, temperature logging, and capacity-retention calculations.
  • If your primary focus is power capability: Require dynamic internal-resistance or impedance measurement synchronized with current and voltage capture across relevant temperatures and states of charge.
  • If your primary focus is early fault detection: Require controlled rest testing that measures self-discharge current or rate and separates true cell leakage from external measurement or balancing currents.
  • If your primary focus is cell-string qualification: Require individual cell voltages, cell-level capacity imbalance, total and localized resistance, and identification of the weakest cell’s self-discharge behavior.
  • If your primary focus is degradation research: Require synchronized raw-data logging of capacity, resistance, self-discharge, temperature, cycle history, voltage limits, C-rate, and exposure time at stressful operating conditions.

A laboratory can evaluate SoH defensibly only when it measures capacity, resistance, and self-discharge under controlled conditions and resolves the weakest cell within every series string.

Summary Table:

Parameter Single Cell Cell String
Deliverable capacity Directly measure usable Ah (C/current) Measure capacity of each cell to find imbalance
Internal resistance Measure via pulse, AC, or EIS Measure total string resistance and localize contributions
Self-discharge Measure current during rest period Identify weakest cell's self-discharge rate
Cell voltage Continuously record voltage Record individual cell voltages (spread)
Charge/discharge current Integrate for capacity and resistance calc Integrate per cell for balancing assessment
Temperature Record cell surface temperature Record temperatures for all cells to ensure uniformity
Cycle count & history Log cycling, DOD, C-rate, and temperature stresses Log cycling and stress history for each cell
Electrochemical impedance Optional for detailed degradation analysis Optional for localized resistance and fault detection

Maximize Battery Performance with KINTEK's Precision Testing Instruments

Take the guesswork out of SoH assessment. KINTEK's advanced battery testing equipment delivers accurate measurements of capacity, internal resistance, and self-discharge for single cells and strings—so you can identify weaknesses early and extend battery life. Our comprehensive portfolio supports your entire cell fabrication workflow, including precision pressing and testing systems. Contact us today for a customized solution: Contact KINTEK!